Note: this product is available in several different variations. Use the dropdown box to make a selection when placing your order.
Work safely in potentially explosive environments with this instrinsically safe pressure calibrator!
Classified with IECEx and Atex Ex iA IIB T3 Gb (zone 1) rating and featuring two isolated, stainless steel pressure sensors with accuracy of 0.025%, the 712Ex ensures the safety of the user and caters for practically any pressure calibration need.
Pressure measurements can be displayed in a total of 19 engineering units, including psi, bar, kPa, kgcm², mmHg and several others (full list available in the attached datasheet). The instrument is suitable for measuring 4 to 20mA signals, has a large backlit graphical display and includes the ability to store five instrument setups for easy recall and use.
Fluke 712Ex Intrinsically Safe Dual Sensor Pressure Calibrator Key Features
Selection of models available - choose a sensor type and pressure range
Intrinsically safe - IECEx and Atex Ex iA IIB T3 Gb (zone 1) rating
Ideal for gas flow calibration (custody transfer) applications
Dual sensor - two isolated, stainless steel pressure sensors with 0.025% accuracy
Selectable measurement units - 19 different options
Pt100 RTD input for temperature measurements (probe optional)
Measures 4 to 20mA signal
Large backlit graphical display shows up to three inputs simultaneously
Important Notice This item is made-to-order exclusively for you once your order is placed. As a result, it is not eligible for cancellation or return under our standard policy. Your statutory rights remain unaffected.
Spend £150+ (Ex VAT) on any qualified Fluke or Beha-Amprobe product(s) and claim a free tool of your choice from the list at Fluke Promotion. Valid February 17 2025 - July 31 2025.
Claims are made direct to Fluke. Terms and conditions apply.
Technical Information
Specification
Fluke 721Ex Technical Specifications
Instrument Setup Recall
Last used on powerup
Operating Temperature
-10 to +45°C
Storage Temperature
-20 to +60°C
Altitude
2000m
Ingress Protection
IP40
Power Requirements
6.0V DC
Battery Life
>35 hours, typical usage
Batteries Required
4 x standard AA cells
Dimensions (H x W x D)
7.9 x 4.3 x 2.3" (20 x 11 x 5.8cm)
Connectors/Ports Pressure
Two, 1/8" NPT, 700P series pressure module; RTD probe
Atex
Ex ia IIB T3 Gb (Ta=-10 ... +45°C) KEMA 10 ATEX 0168X
IECEx
Ex ia IIB T3 Gb (Ta=-10...+45°C) II 2 G IEXEx CSA 10.0013X