Note: there are two versions of this product available: 30G and 100G. Choose an option using the dropdown box above.
No more carrying around a huge pressure calibrator; the Fluke 719 is a handheld instrument with a built-in electric pump, making it easy to perform pressure calibration with one hand!
Highly accurate with best-in class pressure measurement uncertainty of 0.025%, the Fluke 719 also includes programmable pump lmit settings to eliminate over pressurisation. This unit is also easy to clean, with unique cleanout port design to reduce pump failure, and easy cleaning without repair.
The Fluke 719 can measure and source mA with high accuracy of 0.15% and also perform simulataneous pressure measurements, allows simulation of mA signals to troubleshoot 4-20mA loops and can test pressure switches with the built-in switch test function.
The measurement range of this product can be further expanded by using one of 29 Fluke 700Pxx pressure modules (sold separately).
Get a free gift this Black! Even better, this bonus stacks with any other offer.
Free item will automatically add to your basket. Limited Stock. Offer is based on ex VAT basket total after discounts and before shipping costs. See Black Friday Offers for details.
Remember: We will not be beaten on price. Contact us if you've seen a cheaper price.
Buy a qualifying Fluke Industrial, Fluke Networks or Beha-Amprobe product and earn 1 point for every £1 spent. Redeem points for rewards straight away or save for higher-value gifts.
Offer runs 1 September–31 December 2026. Claims are made through Fluke and offer works in conjuction with our other offers.
| Fluke 719-30G | |
| Range | -12 to 36psi (-850mbar to 2.4bar) |
| Resolution | 0.001psi (0.1mbar) |
| Accuracy | 0.025% range for 6 months |
| 0.035% of range for 1 year | |
| Fluke 719-100G | |
| Range | -12 to 120psi (-850 to 8bar) |
| Resolution | 0.01psi (1mbar) |
| Accuracy | 0.025% range for 6 months |
| 0.035% of range for 1 year | |
| mA | |
| Range | 0 to 24mA |
| Resolution | 0.001mA |
| Accuracy | 0.015% Rdg + 2 counts |
| Notes | Max load, 1000Ω (750Ω in HART mode) |
| Loop Supply | |
| Range | 24V DC |
| Resolution | N/A |
| Accuracy | ±10% |
| General Specifications | |
| Storage Temperature | -30 to 60°C |
| Operating Temperature | -10 to 55°C |
| Relative Humidity | 95% -10 to 30°C |
| 75% (30 to 40°C) | |
| 45% (40 to 50°C) | |
| 35% (50 to 55°C) | |
| Shock | 1m drop test |
| Safety | CSA C22.2 No 1010.1: 1992 |
| EMC | EN50082-1: 1992 and EN55022: 1994 Class B |
| Size (H x W x L) | 60 x 87 x 210mm (2.19 x 3.41 x 8.28 inches) |
| Weight | 912g (2lbs) |
| Battery | 2 x 9V alkaline |
| Battery Life | 12 hours with 12mA into 500Ω, 200 pump cycles to 100 psi, 500 pump cycles to 30 psi |
| Warranty | Three years, one year for pump |
| Pump Specifications | |
| Maximum Pressure Generation | 120PSI (8 bar) (100PSI model) |
| Pump To (with 1m of 1.8" test hose) | -30 PSI in less than 15 seconds |
| -100 PSI in less than 45 seconds | |
| Pump Cycles Between Rebuilds | 5000 |
| Total Pump Life | 20000 calibration cycle |