The Fluke 773 is designed to let users measure 4 to 20mA signals without breaking the loop, saving time and effort when carrying out testing procedures.
This device is the best overall process clamp meter available from Fluke and includes a massive selection of advanced features designed to make your time spent testing as simple as possible. In addition to carrying out no-break-loop testing of 4 to 20mA signals this device can also measure 10 to 50mA signals in older control systems by using the 99.99mA range mode and it can also carry out in-circuit measurements, can be used to source signals for testing control systems (I/O and I/Ps) and much more.
This process clamp meter also includes a detachable clamp with extension cable for measurement in difficult, tight locations, can be used to source and measure VDC, can power a transmitter and includes a dual backlit display which shows both mA measurements and percent of 4 to 20mA span.
To aid with testing in darkened areas the Fluke 773 also includes a measurement spotlight for easy spotting hidden wires.
Get a free gift this Black! Even better, this bonus stacks with any other offer.
Free item will automatically add to your basket. Limited Stock. Offer is based on ex VAT basket total after discounts and before shipping costs. See Black Friday Offers for details.
Remember: We will not be beaten on price. Contact us if you've seen a cheaper price.
Buy a qualifying Fluke Industrial, Fluke Networks or Beha-Amprobe product and earn 1 point for every £1 spent. Redeem points for rewards straight away or save for higher-value gifts.
Offer runs 1 September–31 December 2026. Claims are made through Fluke and offer works in conjuction with our other offers.
| Diameter of measurable conductor | 0.177” or 4.5 mm max |
| Operating temperature | -10 to 50 ºC |
| Storage Temperature | -25 to 60ºC |
| Operating Humidity | < 90% @ <30 ºC, < 75% @ 30 to 55ºC |
| Operating Altitude | 0 to 2000 m |
| IP Rating | IP 40 |
| Size | 44 x 70 x 246 mm (2 7/8 x 5 ¾ x 11 5/8 in) |
| Weight | 410 g, (14.4 oz) |
| Vibration | Random 2 g, 5 to 500 Hz |
| Shock | 1 meter drop test (except the jaw) |
| EMI/RFI | Meets EN61326-1 Note: For current measurement w/ JAW, add 1mA to specification for EMC field strengths of 1V/m up to 3V/m. |
| Temperature Coefficient | 0.01% ºC (<18ºC or > 28ºC) |
| Power, battery life | (4) AA 1.5 V Alkaline, IEC LR6, 12 hours in mA source into 500 ohms |
| Warranty | Three years for electronics one year for cable and clamp assembly |
Functional Specifications |
|
| mA measurement (Measured by clamp) 771/772/773 |
Resolution and Range: 0 to 20.99 mA Accuracy: 0.2%+2 counts Resolution and Range: 21.0 mA to 100.0 mA Accuracy: 1% + 5 counts |
| mA measurement (Measured in series with test jacks) 772/773 |
Resolution and Range: 0 to 24.00 mA Accuracy: 0.2% + 2counts |
| mA source (Maximum mA Drive: 24 mA into 1,000 ohms) |
Resolution and Range: 0 to 24.00 mA Accuracy: 0.2% + 2counts |
| mA simulate (Maximum voltage 50 V dc) |
Resolution and Range: 0 to 24.00 mA |
| Voltage source (2 mA maximum drive current) | Resolution and Range: 0 to 10.00 V dc Accuracy: 0.2% + 2counts |