The Seaward Solar PV:1525 supports testing in systems up to 1500V DC and 25A. It provides safety and performance checks, including ground continuity, insulation resistance, open-circuit voltage, and short-circuit current measurements.
The PV:1525 includes Bluetooth connectivity for easy data transfer to a companion mobile app. Simplifying result logging and tagging while enabling wireless firmware updates. Additional highlights include illuminated function buttons for guided testing, onboard memory for storing test records, and a manual diagnostic mode for in-depth troubleshooting.
| Seaward PV:1525 Unit Only | Seaward PV:1525 Kit | |
| Seaward PV:1525 Unit | ✓ | ✓ |
| Set 4mm Test Probes: Red, Black and Yellow (with detachable clips) | ✓ | ✓ |
| Set Instrument to MC4 Test Leads - Yellow and Black | ✓ | ✓ |
| LiPo Battery Pack | ✓ | ✓ |
| Battery Charger | ✓ | ✓ |
| Carry Case | - | ✓ |
| Quick Start Guide | ✓ | ✓ |
| SS200LR Irradiance Meter | - | ✓ |
| PV:1500 Clamp (Wireless Clamp Meter) | - | ✓ |
| Solar Panel Clamp | - | ✓ |
| Seaward SolarCert Software License | - | ✓ |
| UKAS Calibration Certificate | ✓ | ✓ |
| Open Circuit Voltage Measurement (PV Terminals) | |
|---|---|
| Display Range | 0.0VDC - 1500VDC |
| Measuring Range | 5.0VDC - 1500VDC |
| Resolution | 0.1VDC maximum |
| Short Circuit Current Measurement (PV Terminals) | |
| Display Range | 0.00ADC - 25.00ADC |
| Measuring Range | 0.50ADC - 25.00ADC |
| Maximum Power | 37.5kW |
| Resolution | 0.01 ADC maximum |
| IV Curve Measurement (PV:1525-IV Only) | |
| Number of Test Points | Dynamic |
| Max Power | Depending upon Irradiance and Panel Technology |
| Earth Continuity / Resistance Measurement (4mm Terminals) | |
| Test Voltage Open Circuit | >4VDC, nominal |
| Test Current into 2Ω | >200mA |
| Display Range | 0.00Ω - 199Ω |
| Measuring Range (EN 61557-4) | 0.05Ω - 199Ω |
| Resolution | 0.01Ω maximum |
| Test Leads Zero | Zero up to 10Ω |
| Visible Warning | ≥30V AC or DC at inputs |
| Circuitry Protection | Test inhibited if ≥30V AC or DC at inputs |
| Repeat Tests as per IEC61557-4 | Approx 4000 1-second tests |
| Insulation Resistance Measurement (4mm Terminals) | |
| Test Voltage Open Circuit | 250, 500, 1000, 1500V (as per IEC61557-2) |
| Test Current Short Circuit | 2mA |
| Display Range | 0.05MΩ - 300MΩ |
| Measuring Range (EN61557-2) | 0.05MΩ - 300MΩ |
| Resolution | 0.01MΩ Maximum |
| Visible Warning | ≥30V AC or DC at inputs |
| Circuitry Protection | Test inhibited if ≥30V AC or DC at inputs |
| Repeat Tests as per IEC61557-4 | Approx 4000 1-second tests |
| RPE Voltage Measurement (4mm Terminals) | |
| Display Range | 30V - 440VAC/DC |
| Voltage Measuring Range | 30V - 440VDC 30V- 440VAC 50-60Hz |
| Resolution | 1V |
| Operating Current (Wireless via PV:1500 Clamp) | |
| Display Range | 0.1A - 40.0A |
| Current Measuring Range | 0.1A - 40.0A DC 0.1A -40.0A AC 50-60Hz |
| Operating Power (PV Terminals & wireless using PV:1500 Clamp) | |
| Display Range | 0.00kW - 40kW |
| Measuring Range | 0.50kW - 40kW |
| Resolution | 0.01kW |
| General | |
| Weight | 1kg / 2.2lb |
| Dimensions | 265 x 115 x 78 mm 10.4 x 4.6 x 3.1 in |
| Display | 3.5-inch colour display with Backlight |
| Power Source | 11.55V 2930mAh LiPo Battery |
| Auto Power Down | User Configurable |
| Onboard Memory | 1000 Slots |