The PRTs are designed to perform in extreme environments where both the transition junction and the lead wires are required to withstand temperatures covering the entire operating range of the probe. Termination Key.
The probes are just two inches (50 mm) in length with a sheath diameter of 1/8 inch (3.1 mm). Since it can be fully immersed over its entire temperature range, you don't have to worry about calculating minimum immersion depth - immerse the entire probe, transition junction, and lead wires, in either non-corrosive liquids or dry mediums ranging from -200 °C to 160 °C. Termination ranges available from A to S & CLEN A to CLEN P.
Product Ranges
Fluke 5606-50-A Immersion Probe PRT, -200° to 160°, 50mm
Fluke 5606-50-B Immersion Probe PRT, -200° to 160°, 50mm
Fluke 5606-50-D Immersion Probe PRT, -200° to 160°, 50mm
Fluke 5606-50-G Immersion Probe PRT, -200° to 160°, 50mm
Fluke 5606-50-J Immersion Probe PRT, -200° to 160°, 50mm
Fluke 5606-50-L Immersion Probe PRT, -200° to 160°, 50mm
Fluke 5606-50-M Immersion Probe PRT, -200° to 160°, 50mm
Fluke 5606-50-P Immersion Probe PRT, -200° to 160°, 50mm
Fluke 5606-50-S Immersion Probe PRT, -200° to 160°, 50mm
Fluke 5606-CLEN-A Immersion Probe PRT, -200° to 160°, 50mm
Fluke 5606-CLEN-B Immersion Probe PRT, -200° to 160°, 50mm
Fluke 5606-CLEN-D Immersion Probe PRT, -200° to 160°, 50mm
Fluke 5606-CLEN-G Immersion Probe PRT, -200° to 160°, 50mm
Fluke 5606-CLEN-J Immersion Probe PRT, -200° to 160°, 50mm
Fluke 5606-CLEN-L Immersion Probe PRT, -200° to 160°, 50mm
Fluke 5606-CLEN-M Immersion Probe PRT, -200° to 160°, 50mm
Fluke 5606-CLEN-P Immersion Probe PRT, -200° to 160°, 50mm
Termination Key:
A (INFO-CON for 914X)
B (Bare wire)
D (5-pin DIN for Tweener Thermometers)
G (Gold pins)
J (Banana plugs)
L (Mini spade lugs)
M (Mini banana plugs)
P (INFO-CON for 1523 or 1524)
S (Spade lugs)
£572.00
£686.40 inc vat
Free Delivery*
Important Notice This item is made-to-order exclusively for you once your order is placed. As a result, it is not eligible for cancellation or return under our standard policy. Your statutory rights remain unaffected.
Spend £150+ (Ex VAT) on any qualified Fluke or Beha-Amprobe product(s) and claim a free tool of your choice from the list at Fluke Promotion. Valid February 17 2025 - July 31 2025.
Claims are made direct to Fluke. Terms and conditions apply.
Technical Information
Specification
Fluke 5606 Specifications
Temperature range
–200 °C to 160 °C.
Nominal resistance at 0.01 °C
100 Ω ± 0.1 Ω.
Temperature coefficient
0.00385 Ω/Ω/°C.
Accuracy[1]
± 0.04 °C at 0.01 °C ± 0.06 °C at 160 °C.
Short-term repeatability[2]
± 0.03 °C at 0.01 °C ± 0.04 °C at 160 °C.
Drift[3]
± 0.03 °C at 0.01 °C ± 0.04 °C at 160 °C.
Hysteresis
± 0.015 °C.
Sheath length
50 mm ± 5 mm, (2 in ± 0.2 in).
Sheath diameter
3.1 mm ± 0.1 mm (1/8 in ± 0.004 in).
Sheath material
316 SST.
Transition junction temperature range[4]
–200 °C to 160 °C.
Sensor length
30 mm ± 3mm (1.2 in ±0.1 in).
Sensor location
3 mm ± 1 mm from tip (1.2 in ± 0.04 in).
Minimum insulation resistance
20 MΩ at 23 °C.
Minimum immersion length
Full immersion.
Maximum immersion depth in liquid medium
Full immersion.
Response time[5]
12 seconds typical.
Self-heating (in 0 °C bath)
± 0.003 °C.
Lead-wire cable type
Enameled copper wire.
Lead-wire length
2.4 meters (8 ft).
Lead-wire temperature range
160 °C.
[1] Includes calibration accuracy and 100 hr drift (k = 2).
[2] Three thermal cycles from min to max temp, including hysteresis, 95 % confidence (k = 2).
[3] After 100 hours at max temp, 95 % confidence (k = 2).
[4] Temperatures outside this range will cause irreparable damage.