The Hanna Instruments HI-2631-02 measures conductivity, total dissolved solids (TDS), salinity, and temperature. It features built-in data logging, allowing data to be transferred to a USB drive or computer. It utilises the HI-763100 digital conductivity electrode, which features an integrated temperature sensor and automatic probe recognition via a unique serial number.
A Basic Mode simplifies setup and measurement for routine testing, while Standard Operating Mode provides full access to configuration and calibration functions. The instrument has a 5.5-inch LCD with a wide 130° viewing angle and capacitive touch keys for navigation.
Typical applications: laboratories, teaching environments, and industrial conductivity testing.
Get a free gift this Black! Even better, this bonus stacks with any other offer.
Free item will automatically add to your basket. Limited Stock. Offer is based on ex VAT basket total after discounts and before shipping costs. See Black Friday Offers for details.
Remember: We will not be beaten on price. Contact us if you've seen a cheaper price.
| EC- using HI-763100 Four-Ring Conductivity Probe | |
|---|---|
| EC Range | 0.00 to 29.99 µS/cm 30.0 to 299.9 µS/cm 300 to 2999 µS/cm 3.00 to 29.99 mS/cm 30.0 to 200.0 mS/cm up to 500.0 mS/cm and absolute conductivity (TDS) is the conductivity value without temperature compensation |
| Resolution | 0.01 µS/cm 0.1 µS/cm 1 µS/cm 0.01 mS/cm 0.1 mS/cm |
| Accuracy (@25°C) | ±1% of reading (±0.5 µS/cm or 1 digit, whichever is greater) |
| Calibration | Cell factor calibration; six standards available: 84µS/cm 1413 µS/cm 5.00 mS/cm 12.88 mS/cm 80.0 mS/cm 118.8 mS/cm one point offset: 0.00 µS/cm |
| Temperature Compensation | Automatic, -5 to 100.0°C NoTC can be selected to measure absolute conductivity |
| Temperature Coefficient | 0.00 to 6.00%/°C (for EC and TDS only), default value is 1.90%/°C |
| TDS - using HI-763100 Four-Ring Conductivity Probe | |
| TDS Range | 0.00 to 14.99 mg/L (ppm) 15.0 to 149.9 mg/L (ppm) 150 to 1499 mg/L (ppm) 1.50 to 14.99 g/L 15.0 to 100.0 g/L up to 400.0 g/L absolute TDS using 0.80 conversion factor |
| Resolution | 0.01 mg/L (ppm) 0.1 mg/L (ppm) 1 (ppm) 0.01 g/L 0.1 g/L |
| Accuracy (@25°C) | ±1% of reading (±0.03 ppm or 1 digit, whichever is greater) |
| Calibration | Through EC calibration |
| TDS Factor in Standard Mode | 0.40 to 0.80 (default value is 0.50) |
| Salinity - using HI-763100 Four-Ring Conductivity Probe | |
| Salinity Range | 0.0 to 400.0 % NaCl 2.00 to 42.00 PSU 0.00 to 80.00 g/L |
| Resolution | 0.1 % NaCl 0.01 PSU 0.01 g/L |
| Accuracy (@25°C) | ±1% of reading |
| Calibration % NaCl only | One point with the HI-7037 standard |
| Temperature | |
| Temperature Range | -20.0 to 120.0 °C |
| Resolution | 0.1°C |
| Accuracy | ±0.5°C |
| Additional Specifications | |
| Logging | Up to 1000 (standard mode) (400) records organised in: • Manual log-on demand, maximum 200 logs • Manual log-on-stability, maximum 200 logs • Interval logging in Standard Mode, maximum 600 logs (100 lots) |
| PC Interface | USB-C |
| Storage Interface | USB-C |
| Power Supply | USB Type C (5 VDC; 500mA) |
| Environment | 0 to 50˚C Maximum 95% RH non-condensing |
| Dimensions | 205 x 160 x 77 mm |
| Weight | Approximately 0.85 kg |