The Rigel SafeTest 60+ is a medical safety analyser for testing medical and laboratory equipment. It is ideal for devices that don't require patient-led testing, like beds, hoists, infusion pumps, CPAPs, centrifuges, etc.
Additionally, the SafeTest 60+ includes basic testing capabilities for equipment with applied parts, including patient current leakage and insulation measurements.
This safety analyser supports compliance with a variety of international safety standards, including leakage testing to IEC 60601, 62353, 61010, and NFPA-99, earth bond testing to IEC 62353, 61010, and NFPA-99, and insulation testing to IEC 62353.
The SafeTest 60+ offers flexibility with a wide range of power configurations to meet local and international market needs. Its large colour display and colour-coded user interface simplify test selection with a single key press. In contrast, the fast step-through test routine ensures a quick and reliable testing process.
Get a free gift this Black! Even better, this bonus stacks with any other offer.
Free item will automatically add to your basket. Limited Stock. Offer is based on ex VAT basket total after discounts and before shipping costs. See Black Friday Offers for details.
Remember: We will not be beaten on price. Contact us if you've seen a cheaper price.
| Earth Continuity | |
|---|---|
| Pre-pulse | 2-wire ZAP technique |
| Pulse Shape | Exponential Decay |
| Decay Time | 200 - 550μs to 5% of peak (0.1 to 0.8Ω respectively) |
| Method | Wire |
| Measurement Current | >± 200mADC into 2Ω |
| Max Test Voltage | 4-24Vrms o/c |
| Measuring Range (Low Range) | 0.001 – 0.999Ω |
| Resolution | 0.1Ω |
| Measuring Range (Mid Range) | 1.00 – 9.99Ω |
| Measuring Range (High Range) | 10.0 – 19.9Ω |
| Accuracy | ± 1% of value, ± 5mΩ ± 1% of value, ± 5mΩ |
| Circuit Protection | Test inhibited if ≥ 30VAC or DC at 4mm inputs |
| Insulation Resistance | |
| Measurement | Point-to-point and EUT to Earth |
| Voltage | 500VDC @1mA |
| Maximum O/C Voltage | <600VDC |
| Range | 100K – 100MΩ ± 5% ± 2 Digits |
| Resolution | 0.01MΩ |
| Short Circuit Current | <2mA |
| Voltage Measurement | |
| Application | L-N, L-E, N-E and touch voltage (IEC 61010) |
| Range | 0.0V – 300VAC |
| Resolution | 0.1Hz |
| Accuracy | Unspecified |
| Mains frequency | 45.0 – 66.0Hz |
| Powered Leakage Measurements | |
| IEC 62353 | Equipment Leakage (Direct) |
| IEC / AAMI 60601 | Earth + Enclosure Leakage + Patient Leakage |
| NFPA-99 | Ground + Chassis Leakage + Patient Leakage |
| IEC 61010 | Touch Leakage, Touch Voltage |
| Test Voltage | Mains Supply Voltage |
| Measuring Range | 0.1 - 9999μA (0.1 - 8000μA typical for IEC61010) |
| Measurement/Display Resolution | 0.1μA |
| Accuracy | ± 2%, ± 5μA |
| Mains Reversal | Soft Key |
| Single Fault Conditions | Open Neutral and Open Earth Via Soft Key |
| Frequency Response | IEC 60601 – 62353, NFPA-99, and IEC 61010 selectable |
| EUT Load Current Measurement | |
| Range | 0.0A – 20.0A |
| Resolution | 0.1A |
| Accuracy | ±5% ± 2 digits |
| General Specifications | |
| Weight | 1.1kg/2lbs (unit) 2.2kg/5lbs (complete with accessories) |
| Size (L x W x D) | 225 x 150 x 100mm / 9 x 6 x 4” |
| Operating Temperature | 0 to 40°C |
| Humidity | 0 – 98% Relative humidity, non-condensing |
| Storage Temperature | -10 to 50°C |
| Altitude | 0 – 2000m |
| Ingress Protection | IP 40 |
| Operating Pollution Degree | 2, according to IEC 60529 |
| Overvoltage Category | CAT II 300V |
| Power Source | |
| Maximum Current Rating | 20A @ 120V / 16A @ 230V |
| Duty cycle (@21°C ambient) | 16A to 20A, 3 min. on/ 10 min. off 10A to 15A, 3 min. on/ 5 min. off 0A to 10A, continuous |
| Mains Power | 90-264V 48-64Hz |