The Rigel SafeTest 60+ is a medical safety analyser for testing medical and laboratory equipment. It is ideal for devices that don't require patient-led testing, like beds, hoists, infusion pumps, CPAPs, centrifuges, etc.
Additionally, the SafeTest 60+ includes basic testing capabilities for equipment with applied parts, including patient current leakage and insulation measurements.
This safety analyser supports compliance with a variety of international safety standards, including leakage testing to IEC 60601, 62353, 61010, and NFPA-99, earth bond testing to IEC 62353, 61010, and NFPA-99, and insulation testing to IEC 62353.
The SafeTest 60+ offers flexibility with a wide range of power configurations to meet local and international market needs. Its large colour display and colour-coded user interface simplify test selection with a single key press. In contrast, the fast step-through test routine ensures a quick and reliable testing process.
Earth Continuity | |
---|---|
Pre-pulse | 2-wire ZAP technique |
Pulse Shape | Exponential Decay |
Decay Time | 200 - 550μs to 5% of peak (0.1 to 0.8Ω respectively) |
Method | Wire |
Measurement Current | >± 200mADC into 2Ω |
Max Test Voltage | 4-24Vrms o/c |
Measuring Range (Low Range) | 0.001 – 0.999Ω |
Resolution | 0.1Ω |
Measuring Range (Mid Range) | 1.00 – 9.99Ω |
Measuring Range (High Range) | 10.0 – 19.9Ω |
Accuracy | ± 1% of value, ± 5mΩ ± 1% of value, ± 5mΩ |
Circuit Protection | Test inhibited if ≥ 30VAC or DC at 4mm inputs |
Insulation Resistance | |
Measurement | Point-to-point and EUT to Earth |
Voltage | 500VDC @1mA |
Maximum O/C Voltage | <600VDC |
Range | 100K – 100MΩ ± 5% ± 2 Digits |
Resolution | 0.01MΩ |
Short Circuit Current | <2mA |
Voltage Measurement | |
Application | L-N, L-E, N-E and touch voltage (IEC 61010) |
Range | 0.0V – 300VAC |
Resolution | 0.1Hz |
Accuracy | Unspecified |
Mains frequency | 45.0 – 66.0Hz |
Powered Leakage Measurements | |
IEC 62353 | Equipment Leakage (Direct) |
IEC / AAMI 60601 | Earth + Enclosure Leakage + Patient Leakage |
NFPA-99 | Ground + Chassis Leakage + Patient Leakage |
IEC 61010 | Touch Leakage, Touch Voltage |
Test Voltage | Mains Supply Voltage |
Measuring Range | 0.1 - 9999μA (0.1 - 8000μA typical for IEC61010) |
Measurement/Display Resolution | 0.1μA |
Accuracy | ± 2%, ± 5μA |
Mains Reversal | Soft Key |
Single Fault Conditions | Open Neutral and Open Earth Via Soft Key |
Frequency Response | IEC 60601 – 62353, NFPA-99, and IEC 61010 selectable |
EUT Load Current Measurement | |
Range | 0.0A – 20.0A |
Resolution | 0.1A |
Accuracy | ±5% ± 2 digits |
General Specifications | |
Weight | 1.1kg/2lbs (unit) 2.2kg/5lbs (complete with accessories) |
Size (L x W x D) | 225 x 150 x 100mm / 9 x 6 x 4” |
Operating Temperature | 0 to 40°C |
Humidity | 0 – 98% Relative humidity, non-condensing |
Storage Temperature | -10 to 50°C |
Altitude | 0 – 2000m |
Ingress Protection | IP 40 |
Operating Pollution Degree | 2, according to IEC 60529 |
Overvoltage Category | CAT II 300V |
Power Source | |
Maximum Current Rating | 20A @ 120V / 16A @ 230V |
Duty cycle (@21°C ambient) | 16A to 20A, 3 min. on/ 10 min. off 10A to 15A, 3 min. on/ 5 min. off 0A to 10A, continuous |
Mains Power | 90-264V 48-64Hz |