The Time Electronics 5080 is used for calibrating PAT testing devices to maintain accuracy over a time period. It's also suitable for use on insulation/continuity testers.
Housed in its own protective carrying case, the 5080 can either be used in a lab or is ideal for porting out into the field to calibrate instruments on the fly. It's also battery powered, meaning you won't have to find a mains socket and don't have to worry about the effects of stray leakage on testing results.
The calibrator checks a connected instrument's earth bond, insulation, leakage, touch leakage and performs a load test. An LCD display also shows voltage and current levels.
Please note the unit does not come with calibration as standard. If you wish to obtain a calibration certificate, please contact our sales team prior to purchase as this has to be ordered with the unit at an additional cost.
Automatically adds to your basket when the total exceeds £500+ (ex VAT). Limited Stock. More details here.
| EARTH BOND: CALIBRATE LOW VALUE RESISTIVE LOADS | |
| Values/Accuracy | 20mΩ, accuracy 5% 190mΩ, 210mΩ, 950mΩ, 1.050mΩ, 1.08Ω, 2Ω, 10Ω, 18Ω, accuracy 1% |
| Open Circuit Voltage Measurements | Range: 0 to 20V Accuracy: 0.25% of full scale |
| High Current AC Measurements | Range: 0 to 50A AC Accuracy: 0.5% of reading +/- 0.2A |
| Low Current AC Measurements | Range: 0 to 2A AC Accuracy: 0.5% of reading +/- 2mA |
| Low Current DC Measurements | Range: 0 to 2A DC Accuracy: 1% of reading +/- 2mA |
| INSULATION: CALIBRATED HIGH VALUE RESISTIVE LOADS | |
| Values/Accuracy | 95kΩ, 105kΩ, 500kΩ, 950kΩ, 1.05MΩ, 5MΩ, 10MΩ, accuracy 0.1% 20MΩ, 50MΩ, 100MΩ, accuracy 1.0% |
| Open Circuit Voltage Measurements | Range: 0 to 1kV DC Accuracy: 0.25% of full scale |
| Current Measurements | Range: 0 to 20mA DC. Accuracy: 0.25% of full scale |
| LEAKAGE: RESISTIVE LOADS FOR LEAKAGE CURRENT | |
| Nominal Leakage Current | At 230V AC of 0.5, 1, 2, 5, 10, 15mA |
| Mains Voltage Measurements | Range: 0 to 500V AC Accuracy 0.25% of full scale |
| Leakage Current Measurements | Range: 0 to 20mA AC Accuracy 0.25% of full scale |
| TOUCH LEAKAGE: CONSTANT CURRENT SOURCE FOR TOUCH LEAKAGE METERS | |
| Current | 1.9mA AC 50 +/-5 Hz, into 2K2 maximum load resistance |
| Current Measurement | Range: 0 to 2mA AC. Accuracy: 0.25% of full scale |
| LOAD TEST: CONNECTION OF EXTERNAL LOAD VIA IEC CONNECTOR | |
| Mains Voltage Measurement | Range: 0 to 500V AC Accuracy: 0.25% of full scale |
| Load Current Measurement | Range: 0 to 13A AC Accuracy: 0.5% of full scale |
| General Specifications | |
| Power | Internal battery 6V re-chargeable NiCad, >200 hrs between charges. |
| Dimensions | W406 x H175 x D330mm |
| Weight | 5.5kg |
| Country of Origin | UK |