Important Notice
This item is made-to-order exclusively for you once your order is placed. As a result, it is not eligible for cancellation or return under our standard policy. Your statutory rights remain unaffected.
These TSI micromanometers measure and monitor airflow and pressure. The PVM610 measures differential and static pressure and velocity using a pitot tube. It does not include features such as sample statistics, volumetric flow rate, or data logging capabilities. The PVM620 includes sample statistics and a volumetric flow rate. It also has data logging software (LogDat2), a K factor, and the ability to measure actual and standard velocity with a variable time constant.
Feature | PVM610 | PVM620 |
---|---|---|
Differential and static pressure | + | + |
Velocity with pitot tube | + | + |
Calibration Certificate | + | + |
Sample statistics | + | |
Volumetric flow rate | + | |
Actual and standard velocity | + | |
Variable time constant | + | |
LogDat2 data logging software | + | |
K factor | + |
PVM610 | PVM620 | ||
---|---|---|---|
Static/Differential Pressure | Range1 | -28.0 to +28.0 mm Hg, -3735 to +3735 Pa (-15 to +15 in. H2O) | |
Accuracy | ±1% of reading ±1 Pa (±0.01 mm Hg, ±0.005 in. H2O) | ||
Resolution | 0.1 Pa, 0.01 mm Hg (0.001 in. H2O) | ||
Velocity from a Pitot Tube | Range2 | 1.27 to 78.7 m/s (250 to 15,500 ft/min) | |
Accuracy3 | ±1.5% at 10.16 m/s (2,000 ft/min) | ||
Resolution | 0.1 m/s (1 ft/min) | ||
Duct Size | N/A | 2.5 to 1270 cm in increments of 0.1 cm (1 to 500 inches in increments of 0.1 in.) | |
Volumetric Flow Rate | N/A | Actual range is a function of velocity, pressure, duct size, and K factor | |
Instrument Temperature Range (Operating) | 5 to 45°C (40 to 113°F) | ||
Instrument Temperature Range (Storage) | -20 to 60°C (-4 to 140°F) | ||
Data Storage Capabilities | N/A | 12,700+ samples and 100 test IDs | |
Logging Interval | N/A | From 1 second to 1 hour | |
Time Constant | N/A | User selectable | |
External Meter Dimensions | 8.4 cm x 17.8 cm x 4.4 cm (3.3 in. x 7.0 in. x 1.8 in.) | ||
Meter Weight with Batteries | 0.6 lbs (0.27 kg) | ||
Power Requirements | Four AA-size batteries or optional AC adapter | ||
1Overpressure range = 7 psi (190 in. H2O, 360 mmHg, 48 kPa). 2 Pressure velocity measurements are not recommended below 1000 ft/min (5 m/s). 3 Accuracy is a function of converting pressure to velocity. Conversion accuracy improves when actual pressure values increase. |