The Aim-TTi EX354RD Digital Bench Power Supply is a dual-output, benchtop, laboratory PSU with two four-digit meters and switchable remote sensing. It has a maximum power output of 280W, a voltage output range of 0 to 35V, and a current output range of 0 to 4A.
The Aim-TTi EX354RD Digital Bench Power Supply’s outputs are independent, isolated, and equipped with a dedicated On/Off switch. These outputs can be wired in series or parallel to provide up to 70V or up to 8A.
Both of the Aim-TTi EX354RD Digital Bench Power Supply’s outputs are intrinsically short-circuit-proof and are protected against external voltages and reverse currents.
Furthermore, to maintain compliance with the safety regulations of most laboratories, the Aim-TTi EX354RD Digital Bench Power Supply’s output terminals have been designed to accept 4mm safety plugs with rigid insulating sleeves. However, to ensure maximum flexibility, these outputs can also accept fork connectors and bare wires.
Each of the Aim-TTi EX354RD Digital Bench Power Supply’s outputs also includes remote sensing terminals which are essential to maintaining precise regulation at the load and true metering of the load voltage. Remote sense terminals can be activated and deactivated via a dedicated switch: when the switch is set to Local the remote sense terminals are deactivated.
The user can view preset voltages and currents before connecting a load using the Aim-TTi EX354RD Digital Bench Power Supply’s DC On/ Off switch. Each of the Aim-TTi EX354RD Digital Bench Power Supply’s outputs features an independent DC On/Off switch.
Each of the Aim-TTi EX354RD Digital Bench Power Supply’s outputs can be set to operate in Constant Current or Constant Voltage mode with automatic crossover and mode indication.
Voltage levels can be configured quickly, easily, and accurately using the Aim-TTi EX354RD Digital Bench Power Supply’s coarse and fine analogue controls. To enable low currents to be set accurately, the Aim-TTi EX354RD Digital Bench Power Supply includes a logarithmic current control.
Voltage and current data are indicated via the Aim-TTi EX354RD Digital Bench Power Supply’s large, bright, four-digit displays with a fixed 10mV and 1mA resolution. The Aim-TTi EX354RD Digital Bench Power Supply’s fixed resolution prevents misinterpretation of readings and improves precision.
Each of the Aim-TTi EX354RD Digital Bench Power Supply has an independent display.
To ensure excellent line and load regulation, low noise, good transient response, and performance close to that of a PSU with an all-linear design, the Aim-TTi EX354RD Digital Bench Power Supply is equipped with Mixed-Mode Regulation. This combines high-frequency, switch-mode regulation with linear post-regulation.
Finally, the Aim-TTi EX354RD Digital Bench Power Supply utilises convection cooling making it entirely free of fan noise.
The Aim-TTi EX354RD Digital Bench Power Supply is a precise, practical, and powerful dual-output PSU.
EX-R Model Range | |||
Voltage/Current Levels | EX355R | 0 to >35V; 0 to >5A | (175W nominal power) |
EX1810R | 0 to >18V; 0 to >10A | (180W nominal power) | |
EX2020R | 0 to >20V; 0 to >20A | (400W nominal power) | |
EX4210R | 0 to >42V; 0 to >10A | (420W nominal power) | |
EX354RD | 2 x (0 to >35V; 0 to >4A) | (280W nominal power | |
EX354RT | 2 x (0 to >35V; 0 to >4A) plus <1.5 to >5.0V @ 5A | (305W nominal power | |
Output Specifications | |||
Output Setting & Control | Voltage Setting | By coarse and fine controls. | |
Current Setting | By single logarithmic control | ||
Output Mode | Constant voltage or constant current with automatic cross-over. CC indicator lit in constant current mode. |
||
Output Switch | Electronic, non-isolating. Preset voltage and current limit are displayed when the Output is off. On state indicated by LED. | ||
Output Performance | Ripple & Noise | Typically <2mV rms, <10mV pk-pk (EX355R, EX354RD, EX354RT) | |
Typically <2mV rms, <20mV pk-pk (EX1810R, EX2020R, EX4210R) (CV mode, 20MHz bandwidth) | |||
Load Regulation | <0·01% of maximum output for a 90% load change (using remote sense). | ||
Line Regulation | <0·01% of maximum output for 10% line change | ||
Transient Response | <200µs to within 50mV of the setting for a 5% to 95% load change. | ||
Temp. Coefficient | Voltage: typically <100ppm /°C. | ||
Output Protection | Outputs will withstand forward voltages up to 40V (EX355R, EX354RD, & EX354RT); 22V (EX1810R); 25V (EX2020R); 48V (EX4210R). Reverse protection by diode clamp for currents to 3A. |
||
Output Connections | Output Terminals | Universal 4mm safety binding posts on 19mm (0·75”) spacing | |
Note | Terminals can accept fixed shroud 4mm plugs, standard 4mm plugs, fork terminals, and bare wires | ||
Remote Sense | Sense Selection | Voltage sensing is selected as Local or Remote by the front panel switch. | |
Sense Terminals | Sprung loaded screw-less terminals | ||
Meter Specifications | |||
Display Type | Dual 4-digit meters, 14mm (0·56”) LED | ||
Voltage Meter | Resolution | 10mV | |
Accuracy | 0·3% of reading ± 3 digits | ||
Current Meter (EX355R, EX354RD, EX354RT) | Resolution | 1mA | |
Accuracy | 0·5% of reading ± 3 digits | ||
Current Meter (EX1810R, EX2020R, EX4210R) | Resolution | 10mA | |
Accuracy | 0·6% of reading ± 3 digits | ||
Auxiliary Output (EX354RT only) | |||
Voltage | Variable <1.5V to >5V by front panel control. | ||
Meter accuracy | 0.3% ± 4 digits. | ||
Current limit | 5A minimum | ||
Load regulation | <0.5% for 90% load change. | ||
Line regulation | <0.1% for 10% line voltage change | ||
Ripple & Noise | Typically <2mVrms, <10mVpk-pk (20MHz bandwidth) | ||
Output Protection | Output will withstand up to 7V forward voltage. Diode clamp reverse protection for currents up to 3A |
||
Output Terminals | Universal 4mm safety binding posts on 19mm (0·75”) spacing | ||
Status Indication | UNREG lamp | ||
General Specifications | |||
Input | AC Input | EX354RD, EX354RT - 110V to 240V 50/60Hz (universal input); Other Models - 230V AC ± 10% 50/60Hz; 115V to order. Installation Category II |
|
VA Ratings |
EX810R, EX355R - 400VA; EX354RD, EX354RT - 500VA EX2020R, EX4210R - 800VA |
||
Temperature & Environmental | Operating Range | +5ºC to +40ºC, 20% to 80% RH | |
Storage Range | -40ºC to + 70ºC | ||
Environmental | Indoor use at altitudes up to 2000m, Pollution Degree 2 | ||
Cooling | EX1810R, EX355R, EX354RD, EX354RT - fan-less convection cooling; EX2020R, EX4210R - fan assisted using low-noise brushless dc fan. |
||
Safety & EMC | Safety | Complies with EN61010-1 | |
EMC | Complies with EN61326 | ||
Physical | Size | Single output models - 140x160x320 mm (WxHxD). Dual and triple output models - 260x160x320 mm (WxHxD) |
|
Weight | EX810R, EX355R - 3.0kg; EX2020R, EX4210R - 3.6kg EX354RD, EX354RT - 4.3kg |