The Megger PAT250 supports both new testers and experienced professionals, with a rechargeable option. It has a rubber casing and hardened, scratch‑resistant glass. Thanks to its battery‑powered design, the PAT250 is perfectly suited for locations where mains power is unavailable or unsafe, such as construction sites, warehouses, and remote areas. Fully compliant with UK and European electrical safety standards, it gives you the confidence to carry out essential PAT testing quickly and safely wherever the job takes you.
This bundle goes beyond the standard tester by providing everything you need to start testing immediately. Alongside the Megger PAT250 (or PAT250 Rechargeable model), you’ll receive all the essential manufacturer accessories, including continuity and voltage test leads, an extension lead adapter, mains plug test lead, and a protective carry case.
To make this a complete, ready‑to‑use solution, the kit also includes our full PAT Essentials package. You’ll benefit from a soft‑shell carrying case, socket tester, microwave leakage detector, IEC adapter set, and a professional CK Tools insulated screwdriver kit. For training and record‑keeping, the bundle features an eLearning PAT Testing Course with exam and certificate, a PASS PAT Testing Logbook, and a comprehensive selection of PAT labels — including passed, circular passed, cable wrap, and failed labels — totalling 1,500 labels to keep you stocked for a long time.
Inside these bundle packs, you will receive the following standard kit from the manufacturer:
You will also receive all of the following as part of our PAT Essentials kit:
|
Keep all of your equipment safe and in one place with this large-sized carrying case! (appearance of case may vary depending on current stock) |
|
Do you need to check if your sockets are working correctly? Just plug this into a standard UK plug socket, and if there's a problem, the socket tester will let you know about it! |
|
Gain a certificate of competency to carry out PAT testing in just two hours with our online PAT course and exam. Teams of staff may be trained by purchasing the course multiple times; each staff member will be assigned a unique activation code and login. |
|
Keep a record of all of your testing with our logbook. The book includes carbon-copied records, allowing you to keep one for yourself and give the other to customers or keep it for reference. |
|
When PAT testing, it's common practice to check that microwaves are also not emitting high levels of radiation. With our microwave leakage detector, you'll be able to do this testing with ease. |
|
You'll need this adaptor to connect your PAT tester to laptop/PC power leads and other similar equipment which has a cloverleaf-shaped connector. |
|
This adaptor is used to connect to laptop power supplies, portable projectors and other similar equipment. |
|
You'll find several appliances such as game consoles, laptop power supplies, audio/visual equipment and more that use this type of connector - with this adaptor, you can connect them up to your PAT tester. |
|
A handy set of insulated screwdrivers (suitable for live working to 1000V) is suitable for a wide variety of applications. |
|
You'll need these every time an appliance passes a test - just tear one off, fill in the fields (appliance ID, your name, test date and next test due date) and stick it on the appliance when done. |
|
These circular-shaped PAT labels are great for plug tops and any other appliance. The labels also include space for ticking the month/year the next test is due. |
|
These passed labels wrap around a wire/cable from an appliance and are recommended for use with power supplies or on appliances where you can't stick a label. They also include space for writing your appliance ID, name, test date and next test date. |
|
You won't need these as often as your passed labels, but you'll need to apply one of these whenever an appliance fails a test. If it does fail, the item should be taken out of service for repair or replacement |
Automatically adds to your basket when the total exceeds £500+ (ex VAT). Limited Stock. More details here.
| Environmental Condition | |
|---|---|
| Operating Ambient | 20 °C |
| Humidity | 45% rh ± 20 points |
| Continuity Test | |
| Test Voltage | Compliance voltage: +4 V DC -0% / +30% (open circuit) |
| Test Current | Bi-directional +200 mA -0 % + 50 mA (into 2 Ω load) |
| Continuity Accuracy | Resistance: ± 5% ± 3 digits (0 to 19.99 Ω) |
| Resolution | Resistance 10 mΩ |
| Display Range | 0.01 to 19.99 Ω |
| Continuity | test nulling up to 9.99 Ω |
| Test Time | User selectable from 2s to 20s or selected during test to 180s |
| Insulation Test | |
| Insulation Test | 250 V DC -0% +25% open circuit 500 V DC -0% +25% open circuit ≥ 500 V 0% DC across 0.5 MΩ load |
| Short Circuit/charge Current | < 2mA DC |
| Insulation Accuracy | ± 3% ± 10 digits (0 to 19.99 MΩ) |
| Resolution | 0.01 mA |
| Display Range | 0.10 MΩ to 99.99 MΩ |
| Test Duration | User selectable from 2s to 20s or selected during test to 180s |
| Substitute Leakage Test | |
| Leakage Current | Accuracy ± 5% ± 3 digits |
| Test Frequency | Nominal mains frequency 50 Hz |
| Test Voltage | < 50V ac |
| Resolution | 0.01 mA |
| Display Range | 0.10 to 19.99 mA |
| Test Duration | 0.01 mA |
| Display Range | User selectable from 2s to 5s Reading corrected to 230 V AC |
| Differential Leakage Current | |
| Test Voltage | Nominal supply voltage 230 V AC |
| Test Frequency | Nominal mains frequency 50 Hz |
| Test Accuracy | ± 5% ± 3 digits ± 3 uA/A |
| Resolution | 0.01 mA |
| Display Range | 0.10 to 19.99 mA |
| Test Duration | User selectable from 2s to 5s |
| Touch Current Test | |
| Test Voltage | Nominal mains 230 V AC |
| Test Frequency | Nominal mains 50 Hz |
| Test Accuracy | ± 5% ± 3 digits |
| Resolution | 0.01 mA |
| Display Range | 0.10 to 3.99 mA |
| Test Duration | User selectable from 2s to 5s |
| SELV Device Test | |
| Test Voltage | 0 to 300 V AC |
| Test Accuracy | ± 3% ± 3 digits |
| Resolution | 0.1 V AC |
| Display Range | 0.1 to 300 V AC |
| Extension Lead Test | |
| Test Voltage | 5 V |
| Polarity | Lead OK Live/neutral shorted Live/neutral reversed Live/neutral open circuit |
| Extension Portable RCD Test | |
| Test Voltage | Nominal mains 230 V |
| Test Frequency | 50 Hz |
| Test Current Accuracy | +2 % to +9 % (1 x I, 5 x I) |
| Test accuracy | Trip time ± 1 % ± 1 ms |
| Resolution | Trip time 0.01 ms |
| Display Range | 0 to 200 ms (1 x I) 0 to 40 ms (5 x I) |
| Main Supply Test | |
| Frequency Measurement Range | 50 Hz |
| Test Voltage | 40 to 300 V AC |
| Test Accuracy | ± 3% ± 3 digits |
| Resolution | 0.1 V AC |
| Display Range | 40 to 300 V AC |
| Circuit Test (Carried out automatically, not available to the user) |
|
| Test Voltage | 5 V |
| Test Frequency | Nominal mains 50 Hz |
| Test Current | < 100mA short-circuit |
| Safety | |
| PAT250 | Instrument designed to IEC 61010-1: 2010 Test leads designed to IEC 61010-031: 2008 300 V to Earth Category II Mains fuse protection to 250 V rms AC |
| EMC Design to meet IEC 61326-1: 2012 and IEC 61326-2-2: 2005. |
|
| Performance Standard | BS EN 50678:2020 BS EN 50699:2020 |
| User Replaceable Fuse | 13 A fused mains plug (PAT250-UK only) One F 100 mA 250 V 5x20 mm HBC |
| Operating Temperature | 0 °C to +40 °C |
| Storage Temperature | -20 °C to +60 °C |
| Humidity | 90 % RH @ +10 °C to +30 °C 75 % RH @ +10 °C to +60 °C |
| Operating Altitude | up to 2,000 m |
| IP Rating | IP40 (with front cover closed) |
| Mechanical | |
| Battery Life | Over 1000 full Class 1 tests using a 3400 mAh Alkaline battery |
| Battery Type | Eight AA LR6 cells 12 V DC (Alkaline cells) 9.6 V DC (NiMH cells) |
| Weight | PAT250 (instrument) 1300g (45.8 oz) - (shipping) 2795g (98.6 oz) PAT250SRX (instrument) 1300g (45.8 oz) - (shipping) 2975g (104.9 oz) |
| Dimensions | Instrument and case 203 (L) x 148 (W) x 78 mm (H) (8 x 5.7 x 3.2 inches) |