or Call 01642 931 329 to discuss alternatives
These self-contained, battery-powered packages contain a pressure generator, fine pressure control, output measurement capabilities, and data storage. The rugged weatherproof design is styled such that the pressure pump can be operated and test leads connected without compromising the visibility of the large dual-parameter display. The mA step and ramp outputs and a built-in continuity tester extend the capabilities to include the commissioning and maintenance of control loops.
A complete maintenance tool and portable calibration standard for pressure instruments and control loops in hazardous 'Zoned' areas. Certified to BASEEFA and CENELEC standards, the calibrator can reduce response times to breakdowns and emergencies by removing the need for 'Hot Permits' and gas detection equipment. This gives peace of mind to all those responsible for safety within hazardous areas.
A highly accurate and easy-to-use calibrator is only part of the solution for improving overall data quality and working efficiency. With data storage and RS 232 interface, calibration times are reduced and data recording errors are eliminated. It also provides error analysis for field reporting of calibration errors and a pass/fail status. In addition, procedures downloaded from a PC automatically configure the calibrator to pre-defined calibration and test routines. The product is available in 2 or 20-bar gauges.
Key Features
- Dedicated task menu
- Certified to European and North American standards
- Pressure switch testing and leak testing
- Pressure transmitter calibration
- Loop testing and fault finding
- Remote pressure sensors
- Portable documenting
- Self-test routines
- Relied upon in the most extreme conditions
- Highly reliable pneumatic and hydraulic assemblies
Product Range
- Druck DPI615SPC-07G Basic Instrument DPI615SPC/IS Portable Pressure Calibrator Intrinsically Safe, 2 Bar Gauge
- Druck DPI615SPC-13G Basic Instrument DPI615SPC/IS Portable Pressure Calibrator Intrinsically Safe, 20 Bar Gauge
Standard Specification | ||
---|---|---|
Pneumatic Calibrator DPI 610/615PC | Hand Pump | -850 mbar to 20 bar capability |
Volume Adjuster | Fine pressure adjustment | |
Release Valve | Vent and controlled release | |
Pressure Port | G ⅛ (female) | |
Media | Most common gases | |
Hydraulic Calibrator DPI 610/615HC | Priming Pump | M5 (female) feed port |
Shut-off Valve | Open for system priming | |
Screw Press | 0 to 400 bar capability | |
Pressure Port | G ⅛ (female) | |
Media | Demineralised water and most hydraulic oils | |
Indicator DPI 610/615I | Release Valve | Vent and controlled release |
Pressure Port | G ⅛ (female) | |
Media | Most common fluids compatible with stainless steel | |
Special Features | Pressure Units | 25 scale units plus one user-defined |
mA Step | Continuous cycle at 10-sec intervals | |
mA Ramp | Continuous cycle with configurable end values and 60 sec travel time | |
Data Log | Multi-parameter with internal memory for 10,000 values. Variable sample period for data logging or log on key press for calibration results |
|
Snapshot | Paperless notepad. Stores up to 20 complete displays | |
RS232 Computer Interface | DPI 610 Unidirectional for uploading results to a PC DPI 615 Bidirectional for downloading procedures and uploading results |
|
Process Functions | Tare, max/min, filter, flow, % span | |
Language | English, French, German, Italian, Portuguese, and Spanish | |
Power Management | Auto power OFF, battery low indicator, and status on key press | |
Display | Panel | 60 x 60 mm graphic LCD |
Readout | ± 99999 capability, 2 readings per second | |
Environmental | Temperature | Operating: -10° to 50°C Calibrated: -10° to 40°C |
Humidity | 0 to 90%, non-condensing | |
Sealing | IP54 | |
Conformity | EN61010, EN50081-1, EN50082-1, CE marked Intrinsically Safe version: Supplied certified for use in hazardous areas EEx ia IIC T4 certificate Ex99E2002X To EN50020: 1994 and EN50014: 1992 Ex ia IIC T4 certificate pending To CAN/CSA - E79-11-95 and CAN/CSA E79 (Division 1, class 1, groups A, B, C, and D) |
|
Physical | 3 kg, 300 x 170 x 140 mm | |
Power Supply | 6 x 1.5 V C cells, alkaline (up to 65 hrs nominal use at 20°C for the standard version and 30 hrs for the IS version). For rechargeable batteries see Option A (20 hrs nominal use) |
SKU | DPI615SPC-07G/13G |
---|