The Fluke i30s is used to measure AC/DC current and complex waveform currents.
It is not a clamp meter, but is an optional current module that is designed for direct attachment to Fluke oscilloscopes and digital multimeters in order to carry out current measurements up to 30A.
The jaws of this clamp are 19mm in diameter.
Specifications
- Specified current range 30 mA to 30A dc, 30 mA to 20 A ac rms
- Usable current range 5 mA to 30 A dc, 30 mA to 20 A ac rms
- Crest Factor 1.4"
- Output sensitivity 100 mV/A
- Accuracy (at +25 °C) ± 1 % of reading ± 2 mA
- Resolution ± 1 mA
- Load impedance > 100 kΩ
- Conductor position sensitivity ± 1 % relative to center reading
- Frequency range DC to 100 kHz (- 0.5 dB)
- Phase shift below 1 kHz < 2 degrees
- Temperature coefficient ± 0.01 % of reading/°C
- Power supply 9 V Alkaline, MN1604/PP3, 30 hours, low battery indicator
- Working voltage (see Safety Standards section) 300 V ac rms or dc
- Maximum conductor size 19 mm diameter
- Output connection Safety BNC connector, supplied with safety 4 mm adapter
- Output zero Manual adjust via thumbwheel
- Cable length 2 meters
- Operating temperature range 0 °C to +50 °C
- Storage temperature range (with battery removed) -20 °C to +85 °C
- Operating humidity 15 % to 85 % (non-condensing)
- Weight 250 g
- Safety standards BS EN 61010-1: 2001 BS EN 61010-2-032: 2002 BS EN 61010-031: 2002 300 Vrms, Category III, Pollution Degree 2 Use of the probe on uninsulated conductors is limited to 300 V acrms or dc and frequencies below 1 kHz.
- EMC Standards EN 61326: 1998 +A1, A2, & A3
0 Reviews
Overall Rating
Additional Information
SKU | I30S |
---|---|
Manufacturer Part Code | 2584935 |
Model Ref | i30s |