All the essential functions you need for testing flue gas appliances in one low cost, easy-to-use, economical instrument.

Suitable for use with all domestic boilers up to 400kW, the KIGAZ 50 flue gas analyser can be used to calculate a wide variety of parameters including oxygen, carbon monoxide, carbon dioxide, flue gas temperature, ambient temperature, draught pressure, losses, air excess, efficiency and opacity index.

This flue gas analyser has 9 pre-programmed combustible fuel types - when in use, you can select natural gas, propane, butane, coke gas, domestic fuel, heavy fuel, biofuel (5%), pellets (8%) or wood (20%) to cater for use in a wide variety of different domestic boiler applications.

The KIMO KIGAZ 50 includes a number of additional features including automatic zeroing in 30 seconds to main accuracy, a backlight for use in poor lighting and the ability to connect via infrared to KIMO's external printer (sold separately) to easily print out your measurement results when working.

KIMO KIGAZ 50 Flue Gas Analyser Key Features

  • KIMO's entry-level flue gas analyser - ideal for comprehensive checks of domestic boilers up to 400kW
  • All the essential functions you need for checking a boiler's operation
  • 9 selectable combustible fuel types - natural gas, propane, butane, coke gas, domestic fuel, heavy fuel, biofuel (5%), pellets (8%) or wood (20%)
  • Fixed oxygen (O2) and carbon monoxide (CO) sensors
  • Measure oxygen from 0 to 21%
  • Measure carbon monoxide from 0 to 8000ppm
  • Calculate carbon dioxide from 0 to 99% volume
  • Flue gas temperature measurements from -100 to +1250°C
  • Ambient temperature measurements from -20 to +120°C
  • Differential pressure (draft) from -20,000 to +20000Pa
  • Calculate losses from 0 to 100%
  • Calculate excess air from 1 to 9.99
  • Calculate lower efficiency from 0 to 100%
  • Calculate higher efficiency (condensing) from 0 to 120%
  • Autozeroing within 30 seconds
  • 10 hour battery life (lithium ion rechargeable)
  • Backlight on the screen for use in poor lighting
  • Output via infrared to external printer (sold separately)

What's Included?

  • KIMO KIGAZ 50 Flue Gas Analyser
  • Transport Bag
  • 180mm Flue Gas Probe with Water Trap
  • USB Cable
  • Mains Adapter
  • Adjustment Certificate

Optional Accessories

  • SKCLD 150 Thermocouple Probe
  • KPD-15 Differential Pressure Kit
  • KEG Gas Network Tightness Kit
  • PMO Opacity Pump
  • CPAK Magnetic Protective Cover
  • KDIP External Printer
Technical Specs

Part Code: KIGAZ 50

KIMO KIGAZ 50 Technical Specifications

ParameterSensorMeasuring RangeResolutionAccuracyT90 Response Time
Oxygen (O2) Electro-chemical From 0 to 21% 0.1% vol. ±0.2% vol. 30s
Carbon Monoxide (CO) Electro-chemical From 0 to 8000ppm 1ppm From 0 to 200ppm: ±10ppm
From 201 to 2000ppm: ±5% of the measured value
From 2001 to 8000ppm: ±10% of the measured value
30s
Carbon Dioxide (CO2) Calculated From 0 to 99% vol. 0.1% vol. - -
Flue Gas Temperature K Thermocouple From -100 to +1250°C 0.1°C ±0.4% (measured value) or ±1.1°C 45s
Ambient Temperature Internal NTC From -20 to +120°C 0.1°C ±0.5°C -
Differential Pressure Draft Semiconductor From -20,000 to +20,000Pa 1Pa From -20000 to -751Pa: ±(0.5% of measured value +4.5Pa)
From 750 to -61Pa: ±(0.9% of measured value + 1.5Pa)
From -60 to 60Pa: ±2Pa
From 61 to 750Pa: ±(0.9% of measured value + 1.5Pa)
From 751 to 20,000Pa: ±(0.5% of measured value + 4.5Pa)
 
Losses Calculated From 0 to 100% 0.1% - -
Excess Air Calculated From 1 to 99.9 0.01 - -
Lower Efficiency Calculated From 0 to 100% 0.1% - -
Higher Efficiency (Condensing) Calculated From 0 to 120% 0.1% - -
Other Specifications
Dimensions Instrument: 240 x 110 x 80mm
Flue gas probe: 180mm
Weight Battery Included: 680g
Display Customised screen
Active view dimensions: 54 x 50mm
Keypad 10 keys dome switch keypad
Material Housing: ABS
Probe Cable: neoprene
Probe: PA5.6.30 GF
Communication Infrared (IrDA technology) between instrument and printer
Power Supply Lithium ion 3.6V 5.2Ah battery
Battery Life 10 hours (continuous operation)
Battery Charging Time 10h with charger and mini-USB cable
Use and Storage Temperature Operating: +5 to 50°C
Storage: -20 to +50°C